• INTRODUCTION
    • In total hip endoprosthetics and consequently for TDA, metal-on-metal combinations are used with the aim of reducing wear debris. In metal-on-metal TDA the release of metal ions has until now been secondary to the main discussion.
  • MATERIALS AND METHODS
    • In order to investigate the ion release following the implantation of the metal-on-metal Maverick type artificial lumbar disc we measured the serum cobalt and chromium concentration following implantation of 15 Maverick TDAs (monosegmental L5/S1, n = 5; bisegmental L4/5 and L5/S1, n = 5; average age 36.5 years). Five healthy subjects (no metal implants) acted as a control group. The two measurements of the metals were carried out using the absorption spectrometry after an average of 14.8 and 36.7 months.
  • RESULTS
    • In summary, the concentrations of cobalt and chromium ions in the serum at both follow-ups amounted on average to 3.3 microg/l (SD 2.6) for cobalt and 2.2 microg/l (SD 1.5) for chromium. These figures are similar to the figures shown in the literature following the implantation of metal-on-metal THA. After a comparison to the control group, both the chromium and cobalt levels in the serum showed visible increases regarding the first and the second follow-up.
  • DISCUSSION
    • As there is still a significant release of cobalt and chromium into the serum after an average follow-up of 36.7 months a persistent release of these ions must be taken into consideration. Despite the evaluation of the systemic and local effects of the release of Cr/Co from orthopaedic implants has not yet been concluded, one should take into consideration an explanation given to patients scheduled for the implantation of a metal-on-metal TDA about these results and the benefits/risks of alternative combinations of gliding contact surfaces.