• BACKGROUND
    • The Lapidus bunionectomy is a popular procedure for severe bunion deformity where metatarsus primus varus is equal to or exceeds 15 degrees. We evaluated a new locking compression plate which may improve outcomes with the Lapidus procedure.
  • METHODS
    • Ten matched pairs of cadaver feet were used to compare the standard crossed 4.0-mm compression screw method of fixation to the LPS Lapidus plate. After performing the matched operations the cadaver constructs were stressed to failure using the INSTRON and Wavemaker software.
  • RESULTS
    • The LPS Lapidus plate load to failure was 108 Nm with a bending moment of 6.0 Nm. The crossed screw technique was inferior at 78 Nm with a bending moment of 4.4 Nm (p = 0.02)
  • CONCLUSION
    • Unlike other H-plates or locking plates, load to failure was higher with the Lapidus plate constructs.
  • CLINICAL RELEVANCE