• BACKGROUND
    • Many surgeons are reluctant to use a constrained liner at the time of acetabular component revision because of concerns that doing so might result in early acetabular component loosening related to high bone-implant interface stresses transmitted from the constrained liner. We hypothesized that, with appropriate initial implant stabilization using highly porous acetabular components with robust supplemental screw fixation, constrained liners could be safely used at the time of acetabular revision.
  • METHODS
    • We retrospectively identified 148 revision total hip arthroplasties (THAs) in which a constrained liner of 1 design was cemented into a newly placed highly porous acetabular component fixed with supplemental screws (mean, 5 screws). The mean age at the time of revision THA was 69 years, and 68% of the patients were female. The most common indications for revision were 2-stage reimplantation (33%), recurrent dislocation (30%), and aseptic loosening of the acetabular component (22%) or both the acetabular and femoral components (9%). The mean duration of follow-up was 7 years.
  • RESULTS
    • There were no failures at the bone-implant interface. No cases of acetabular metal shell loosening were identified on radiographic analysis. The 10-year survival rates with acetabular revision and any reoperation as the end points were 75% and 67%, respectively. Overall, 33 hips (22%) required revision or reoperation because of infection or wound complications (n = 12), dislocation or mechanical failure of the constrained liner (n = 11), periprosthetic femoral fracture (n = 4), femoral loosening (n = 3), or other reasons (n = 3). The 10-year survival rate with re-revision for instability as the end point was 88% overall, which was similar to the 85% 10-year survival rate with re-revision for instability as the end point among hips that were specifically revised because of instability during index revision THA (p = 0.9).
  • CONCLUSIONS
    • Implanting a constrained liner at the time of acetabular revision in high-risk patients resulted in no cases of aseptic metal acetabular component loosening in this large series. This finding is likely related to the fact that a highly porous acetabular component was utilized with robust supplemental screw fixation in each case. Such information is valuable as these data favor a paradigm shift when compared with some traditionally held tenets.
  • LEVEL OF EVIDENCE
    • Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.