• PURPOSE
    • The use of dual mobility cups in total hip arthroplasty has gained popularity in light of the fact it enables to reduce dislocation through increased jumping distance (JD) and impingement-free arc of movement. Modular Dual Mobility Cup (modular DMC) systems have been recently introduced to enable the use of dual mobility cups with standard metal-backed shells. The objective of this study was twofold: calculate the JD for each modular DMC system and conduct a systematic literature review to report clinical outcomes and reasons for failure of this construct.
  • METHODS
    • The JD was calculated using the Sariali formula: JD = 2Rsin [(π/2 - Ψ - arcsin (offset/R))/2]. A qualitative systematic literature review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A comprehensive search for English and French articles between January 2000 and July 2020 was run on PubMed, EMBASE, Google Scholar, and Scopus with the primary objective of finding articles about modular DMC systems.
  • RESULTS
    • We identified eight 8 different manufacturers of modular DMC systems and 327 publications on the subject. After screening for duplicates and eligibility, we identified 229 publications: 206 articles were excluded because they contained no reports on modular DMC systems, whereas other three were not included because they focused on biomechanical aspects. Among the 11 included articles, 2 were prospective case series, 9 were retrospective case series. True dislocation occurred in 25 cases (0.9%), and six of them were solved by closed reduction without necessity of revision, while all 5 intraprosthetic dislocations were operated.
  • CONCLUSIONS
    • Modular DMCs are a valid method to deal with complex THA instability, with good clinical and patient-reported outcomes, low complication rates, and low revision rates at early follow-up. We would advise cautious optimism on the role of modular DMC implants, as it seems safer to use ceramic instead of metallic heads whenever possible to avoid the increase cobalt and chromium trace ion serum levels.