• ABSTRACT
    • Accurate pre-operative templating of prosthesis components is an essential factor in successful total hip arthroplasty (THA), including robotically-assisted THA (RA-THA) techniques. We sought to validate the accuracy of a novel, robotic-optimized THA planning software compared to a predicate THA planner for component sizing. We analyzed a series of 199 patients who received manual THA (mTHA) and fluoroscopy-based RA-THA at a single institution. All cases were templated using a predicate pre-operative templating software. For RA-THA cases, the novel robotic-optimized pre-operative planner software was also used for templating. The differences between templated and implanted acetabular cup, femoral head, and stem component sizes were compared based on matching within 1, 2, and ≥3 sizes. Differences in templated and implanted femoral stem implant geometry were also compared. The robot-optimized pre-operative RA-THA plans demonstrated equivalent accuracy to that of predicate pre-operative plans for both RA-THA and mTHA cases. Templated acetabular cups (90.4 vs. 86.8 vs. 82.8; p = 0.421), femoral stems (76.0 vs. 65.1 vs. 67.7; p = 0.096), and femoral heads (91.3 vs. 96.2 vs. 88.2; p = 0.302) were within +/-1 size of implanted components. No significant differences were detected in the proportion of matching templated and implanted stem geometry across the study cohorts.