• BACKGROUND
    • The MoPyc radial head arthroplasty (RHA) is a monopolar implant with a pyrocarbon head that obtains rigid fixation via controlled expansion of the titanium stem. The aim of this study was to evaluate the short-term to midterm outcomes of MoPyc RHA.
  • MATERIALS AND METHODS
    • Between 2002 and 2021, 139 MoPyc RHA were implanted in 139 patients with a RH fracture. The mean follow-up was 5.9 years ± 3.5 (range 1-16). Range of motion, mean Mayo Elbow Performance Score (MEPS), Quick Disabilities of the Arm, Shoulder and Hand score, visual analog scale (VAS), radiographic outcome, and reason for failure were recorded.
  • RESULTS
    • The mean MEP, Quick Disabilities of the Arm, Shoulder and Hand, and VAS scores were 89.1 ± 2.2 (range, 45-100), 17.5 ± 16.7 (range, 0-78), and 0.8 ± 1.6 (range, 0-50), respectively. Stress shielding (SS) and osteolysis around the stem were identified in 92 (66%) and 20 (14%) patients. A total of 47 (29%) patients experienced at least 1 complication and 29 (21%) of them required re-intervention. Persistent stiffness (n = 12; 9%) was the most common complication. No painful loosening was noted. Osteolysis around the stem, presence of an autoexpanding stem, and overstuffing were associated with a lower MEPS and an increase in VAS (P < .05). SS was associated with an increase in MEPS (adjusted beta coefficients= 6.92; P < .001) and lower VAS (adjusted beta coefficients= -0.69; P = .016). The autoexpending stem increased the likelihood of SS after RHA (adjusted odds ratio = 1.49; P = .001).
  • CONCLUSIONS
    • A well-fixed MoPyc RHA provided satisfactory short to midterm outcomes, without painful loosening. However, the autoexpanding stem system was associated with poorer functional outcomes and increased the likelihood of SS.