Please confirm topic selection

Are you sure you want to trigger topic in your Anconeus AI algorithm?

Please confirm action

You are done for today with this topic.

Would you like to start learning session with this topic items scheduled for future?

Updated: Dec 11 2024

TKA Polyethylene Wear & Manufacturing

Images
https://upload.orthobullets.com/topic/5002/images/poly_moved.jpg
https://upload.orthobullets.com/topic/5002/images/subsurface line_moved.jpg
  • Summary
    • TKA polyethylene wear refers to macroscopic premature failure of polyethylene (PE) due to excessive loading and mechanical loosening
    • Diagnosis is generally made with plain radiographs of the knee showing narrowing of the tibiofemoral implant interface
    • Treatment generally involves revision TKA or isolated polyethylene exchange depending on the stability of the femoral and tibial implants
  • Epidemiology
    • Incidence
      • catastrophic failure is most commonly seen in TKA
        • in contrast to osteolytic failure that is usually seen in THA
        • catastrophic failure may occur in TSA and THA replacement, but less common
  • Etiology
    • Pathophysiology
      • primary variables that lead to catastrophic wear include:
        • PE thickness
        • articular surface design and geometry
          • flat PE should be avoided as knee loads exceed yield strength of UHMWPE in a flat design
          • goals of PE design:
            • maximize contact area
            • minimize contact loads (force/area)
            • biplanar congruency is the best design
              • congruent in both coronal and sagittal planes
        • kinematics
          • sliding wear is bad for PE
            • occurs when ACL is sacrificed and PCL remains
              • most pronounced with CR knee design with a flat PE insert
              • least pronounced with PS or AS knee design with a congruent PE insert
        • PE sterilization
        • PE manufacturing
        • surgical technique
          • tight flexion gap hastens sliding wear effect
          • tight PCL and anterior tibial slope amplify stress
  • Polyethylene thickness
    • Introduction
      • PE insert thickness can be variable depending on manufacturer definition (i.e. some may list PE thickness as the combined thickness of the insert + tibial tray)
        • PE insert labeled as 8 mm, may only have a "true" PE thickness of 4-5 mm at the thinnest point with a ~3 mm thick metal tray
    • Cause of failure
      • PE thickness <8 mm
        • leads to loads transmitted to localized area of PE that exceed PE's inherent yield strength (12-20 mPA)
        • thickness of <8 mm associated with catastrophic PE failure
        • data based on older studies/PE generations, may not be as applicable with modern manufacturing
    • Solution
      • maintain thinnest portion of PE >8 mm
        • a more aggressive tibial cut may avoid having to use a PE insert of <8 mm
          • in younger patients, increased activity combined with thinner PE will increase risk of catastrophic failure
  • Articular surface design and geometry
    • Introduction
      • two general designs in total knee prosthesis include:
        • a deeper congruous joint (deeper cut PE) without rollback
          • less anatomic
          • maximizes contact loads
          • decreases contact stress
        • a flat tibial PE that improves femoral rollback and optimizes flexion
          • more anatomic
          • PCL sparing
          • increases contact stress and risk of catastrophic failure
    • Cause of failure
      • flat designs of tibia PE
        • low contact surface area leads to high contact stress loads in areas of contact
    • Solution
      • increase congruency of articular design
        • higher contact surface area leads to lower contact stress load
        • newer prosthesis designs sacrifice rollback and have a more congruent ("dished") fit between the femoral condyle and the tibial insert in both the sagittal and coronal planes to decrease the contact stress
  • Kinematics
    • Introduction
      • variables that affect kinetics include
        • knee alignment
          • varus alignment of knee associated with catastrophic PE failure
        • femoral rollback
          • optimizes flexion at the cost of increasing contact stress and increased risk of catastrophic failure
    • Cause of failure
      • excessive femoral rollback
        • dyskinetic sliding movements of femur on tibia causes surface cracking and wear
    • Solution
      • perform adequate bone cuts and/or releases to avoid varus malalignment
      • decrease contact stress by minimizing femoral rollback
        • use a more congruous joint design
        • increase posterior slope of tibia
        • use PCL substituting knee for incompetent PCL or dyskinetic femoral rollback
        • to compensate for the lack of rollback, newer designs move the point of contact (where femoral condyle rests) more posterior and have a steeper posterior slope to aid with flexion
  • Polyethylene sterilization
    • Radiation
      • gamma radiation is the most common form of polyethylene sterilization
        • results in oxidized PE that wears poorly and results in osteolysis
      • oxidation vs. cross-linking
        • presence of oxygen determines pathway following free radical formation
          • oxygen rich environment
            • PE becomes oxidized
              • leads to early failure due to
                • subsurface delamination
                • pitting
                • fatigue strength/cracking
          • oxygen depleted environment
            • PE becomes cross-linked
              • improved resistance to adhesive and abrasive wear
              • decrease in mechanical properties (decreased ductility and fatigue resistance)
              • greater risk of catastrophic failure under high loads
            • methods to obtain
              • packing via argon, nitrogen
              • packing in vacuum environment
      • removal of free radicals
        • thermal stabilization/remelting
          • removes free radicals formed during the radiation sterilization process for cross-linking
            • most effective means of removing free radicals as it occurs above the PE melting point
          • changes the PE from its partial crystalline state to its amorphous state
            • disadvantage is that it reduces the mechanical properties of the material
        • annealing
          • maintains its mechanical property
          • less effective at removing free radicals as it occurs below the PE melting point
            • leaves the PE more susceptible to oxidation
    • Solution
      • irradiate PE in inert gas or vacuum to minimize oxidation
  • Polyethylene manufacturing
    • Introduction
      • cutting tools can disrupt chemical bonds of PE
    • Fabrication methods
      • ram bar extrusion and machining
        • UHMWPE powder fed into heated chamber, ram pushes powder into heated cylinder barrel forming a cylindrical rod, cut into 10 ft lengths for sale
        • implants are machined from the cylindrical bar stock
        • leads to variations in PE quality within the bar
      • calcium stearate additive
        • leads to fusion defects in PE
      • sheet compression molding
        • UHMWPE powder introduced into large 4' x 8' rectangular container to make sheets up to 8" thick
        • implants are machined from these molded sheets
      • direct compression molding/net shape
        • best PE fabrication process
        • UHMWPE powder placed into a mold the shape of the final component, which is heated
        • the net shape implant is removed and packaged
        • no external machining involved, implants have high gloss surface finish
        • lower wear rates (50% wear rate of machined products)
          • slow, expensive
    • Cause of failure
      • machining shear forces cause subsurface region (1-2 mm) stretching of PE chains
        • especially in amorphous regions > crystalline regions
      • PE chains are more susceptible to radiation resulting in greater oxidation in this region
        • leads to subsurface delamination and fatigue cracking
          • can show classic white band of oxidation in subsurface (1-2 mm below articular surface) 
      • "Perfect storm" scenario for catastrophic wear
        • metal-backed tibial baseplate with bone-conserving tibial bone cut (thin PE)
        • flat bearing design in coronal plane (low contact area with high contact load)
        • PCL retention with flat PE insert (high sliding wear)
        • ram bar PE with calcium stearate additive (fusion defects in PE)
        • gamma radiation sterilization in air (weakened mechanical properties of PE)
        • machined PE surface (cutting tool stretch effect on the PE)
    • Solution
      • use direct compression molding of PE
        • performed by molding directly from PE powder to the desired product
        • results in less fatigue crack formation and propagation compared to ram bar extrusion
      • avoid machining the articular surface
Card
1 of 4
Question
1 of 23
Private Note